Hub SN-25 Trailer Sheet 25 MAR 2020 - Begin work on Hub SN25 per instructions at /hub/hardware/details/Hub_Inventory_Status_Final_Assembly/hub_msu_final_assembly_prod_only_print.txt - Check GND/ESD continuity - Verify power supply resistances to GND - Preliminary mechanical work - mounting blocks, removing runners, etc. - Begin component R&R. 01 APR 2020 - Completed final assembly of Hub SN25: - Work conformed to given instructions. - Installed RC filter components on MGT_AVCC, MGT_AVTT, and Fan_1V8 supplies. - Installed discrete power wires. - Installed light pipes and front panel. - Installed IPMC module #135 - Powered on and configured DC/DC converters. - Programmed with safe/foundation firmware. - Installed minipods. - Installed MPOs. - Installed FPGA heat sink. - Installed fiber ribbons and clipped to side of heat sink. - Verified 25MHz, 40MHz, and 320MHz clocks with oscilloscope: PLL Free Run Hi-Lock Lo-Lock ------ -------- ------- ------- 40MHz 40.09 40.0843 40.0735 320MHz 320.0 40.0820 40.0760 - Final voltage measurements as follows: Power Bus: FPGA MGT MGT Swch Bulk Fan Bulk Core AVCC AVTT 1V2 1V8 1V8 3V3 ---- ---- ---- ---- ---- ---- ---- Address: 40 41 42 43 44 45 46 Volts: 0.95 1.00 1.20 1.19 1.79 1.82 3.307 Amps: 1.19 0.00 0.06 1.81 0.62 12.31 1.562 Fluke Volts: 0.952 0.999 1.203 1.200 1.799 1.793 3.313 - Removed IPMC bypass (yellow wire) after testing. IPMC module now has direct control of ISO_12V supply. - Unit has been placed in the storage cabinet to wait for its turn in the 14SS. 08 MAY 2020 - Updated this Hub's IPMC firmware to latest version. 05 AUG 2020 - See l1calo/hub/hardware/details/Hub_Inventory_Status_Final_Assembly/FluxProblemOnHubs.txt for details. This note is to confirm that this Hub has been cleaned with G3 flux remover. 10 SEP 2020 - This note is to confirm that this hub has been through the component replacement process described at the bottom of the file at the above link. 24 SEP 2020 - This Hub has been mated with ROD V4 SN10. 11 MAY 2021 - This is a note to confirm that the board has been through the fiber ribbon replacement. See hub/hardware/details/Hub_Inventory_Status_Final_Assembly/Fiber%20Ribbon%20Problem/00_Hub_Fiber_Ribbon_Issue.txt for details.