QC 6 VERSION 5.4: 06-Mar-01 10:33:37

LIST OF FIBERS/TILES WITH RESPONSE < 20% FOR ANL021

SCAN A

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

             

3

10

D4-O

1

1

-22.0

Special Tiles

5

8

C10-O

1

12

-24.1

6

7

C10-E

1

12

-30.7

6

8

C10-E

1

12

-25.5

6

9

C10-E

1

12

-22.1

7

1

A12-O

6

23

-36.9

Fiber replaced

7

3

A12-O

6

23

-39.7

9

5

B11-O

9

26

-21.8

Fiber not changed

9

7

B11-O

9

26

-21.6

12

1

A13-E

24

50*

-23.2

Gap - no problem found

12

2

A13-E

11

37

-20.2

No problem found

15

6

B12-O

5

38

-39.7

On side of tile - fixed

17

10

D5-O

21

38

-20.3

No problem found

37

8

D6-O

15

97

-51.2

Fiber replaced

37

9

D6-O

37

119

-22.3

Fiber replaced

37

10

D6-O

15

97

-46.9

Fiber replaced

37

10

D6-O

74

156

-28.1

Tile replaced

37

10

D6-O

75

157

-30.1

Tile replaced

37

11

D6-O

37

119

-22.3

Fiber replaced

38

10

D6-E

74

156

-32.0

Tile replaced

38

10

D6-E

75

157

-36.2

Tile replaced

41

2

A16-O

46

155

-31.6

Fiber replaced

41

3

A16-O

5

114*

-25.5

Fiber was detached, reglued & added shim

42

3

A16-E

42

151

-21.8

No problem found

43

4

B15-O

16

138

-28.0

Tile replaced, pack 228

QC 6 VERSION 5.4: 08-Mar-01 10:55:19

LIST OF FIBERS/TILES WITH RESPONSE < 25% FOR ANL021

SCAN B

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

             

6

7

C10-E

1

12

-30.7

Special tile

37

8

D6-O

12

94

-42.5

This fiber had been replaced by mistake

37

10

D6-O

12

94

-37.8

41

1

A16-O

41

150

-26.8

This is a new problem - caused by trying to fix something else - fiber not replaced

41

3

A16-O

41

150

-25.9