QC 6 VERSION 5.8: 28-Aug-01 13:13:05

LIST OF FIBERS/TILES WITH RESPONSE < 80%

FOR SCAN A, ANL037

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

             

5

9

C10-O

1

12

-20.3

Special tiles

6

7

C10-E

1

12

-27.7

6

8

C10-E

1

12

-29.5

6

9

C10-E

1

12

-23.1

8

3

A12-E

8

25

-26.0

Reseated profile

9

7

B11-O

1

18*

-22.0

Reseated profile

10

5

B11-E

1

18*

-40.6

Bad slot, fiber not centered on tiles - add shimms

10

7

B11-E

1

18*

-22.8

12

3

A13-E

17

43

-43.5

Loose profile? -reseated

12

3

A13-E

21

47

-49.3

On side of tile - fixed

15

6

B12-O

23

56

-23.3

Loose profile-eseated

15

6

B12-O

24

57

-22.2

Loose profile-eseated

16

5

B12-E

6

39

-47.5

On side of tile - fixed

16

7

B12-E

8

41

-47.6

Loose tile, shimmed tile

16

7

B12-E

10

43

-28.8

Reseated profile

16

7

B12-E

13

46

-43.8

Loose tile - shimmed

17

9

D5-O

38

55

-42.8

On side of tile - fixed

17

9

D5-O

60

77

-20.5

Reseated profile

17

10

D5-O

39

56

-23.4

Reseated profile

18

11

D5-E

17

34*

-56.5

On side of tile - fixed

21

2

A14-O

6

57

-30.8

Loose profile - fixed

22

1

A14-E

15

66*

-20.7

Loose profile - fixed

22

1

A14-E

23

74

-34.9

On side of tile - fixed

22

1

A14-E

25

76

-37.5

Replaced fiber

22

3

A14-E

25

76

-26.7

22

1

A14-E

26

77

-28.9

Replaced fiber

22

3

A14-E

26

77

-25.9

22

3

A14-E

10

61

-59.7

On side of tile - fixed

22

3

A14-E

16

67

-41.8

On side of tile - fixed

22

3

A14-E

20

71

-55.1

On side of tile - fixed

23

5

B13-O

12

72

-23.1

Replaced fiber

23

7

B13-O

12

72

-23.6

24

7

B13-E

9

69

-37.3

On side of tile - fixed

24

7

B13-E

21

81

-21.0

Reseated profile

30

3

A15-E

2

81

-20.6

Added shims

30

3

A15-E

18

97

-60.3

On side of tile - fixed

34

7

B14-E

5

95

-22.5

Reseated profile

34

7

B14-E

26

116

-29.3

Paint in slot - cleaned

37

9

D6-O

73

155

-24.9

Loose profile - reseated

37

11

D6-O

68

150

-26.8

On side of tile - fixed

38

8

D6-E

72

154

-21.2

Shimmed , reseated

38

9

D6-E

36

118

-34.0

On side of tile - fixed

38

9

D6-E

37

119

-52.0

Replaced fiber

38

11

D6-E

37

119

-53.8

38

10

D6-E

75

157

-20.9

Reseated profile

42

1

A16-E

32

141

-20.7

Reseated profile

42

1

A16-E

33

142*

-81.9

On side of tile - fixed

42

3

A16-E

33

142*

-75.8

On side of tile - fixed

42

1

A16-E

48

157

-59.3

On side of tile - fixed

42

2

A16-E

8

117

-30.9

Replaced fiber

42

2

A16-E

12

121

-26.8

On side?, shimmed

42

2

A16-E

16

125

-20.8

Shimmed, reseated

42

2

A16-E

21

130*

-43.0

Replaced fiber

42

2

A16-E

22

131

-29.3

Shimmed

42

3

A16-E

21

130*

-61.8

On side of tile - fixed

42

3

A16-E

34

143

-30.0

On side of tile - fixed

42

3

A16-E

37

146

MISSING

On side of tile - fixed

42

3

A16-E

43

152

-54.5

On side of tile - fixed

44

4

B15-E

10

132

MISSING

Replaced fiber

44

6

B15-E

10

132

MISSING

44

7

B15-E

20

142*

MISSING

On side of tile - fixed

NOTE: most of the problems with fibers on the side of tiles were due to extra wide slots - these usually required adding some shims beside the tiles to center them in the slot.

 

QC 6 VERSION 5.8: 6-Sep-01 17:17:00

LIST OF FIBERS/TILES WITH RESPONSE < 80%

FOR SCAN B of ANL037

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

             

3

11

D4-O

13

13

-22.8

New, profile reseated

3

11

D4-O

15

15

-39.5

On edge of tile - reseated

5

7

C10-O

1

12

-21.3

Special tiles

6

7

C10-E

1

12

-46.4

6

8

C10-E

1

12

-30.6

6

9

C10-E

1

12

-22.5

10

7

B11-E

1

18*

-23.0

Bad slot, shimmed

12

1

A13-E

24

50*

-29.6

Loose profile, taped

15

6

B12-O

25

58

-20.7

New, shimmed tile

17

10

D5-O

43

60

-21.2

New, probably tile, no packs with better light available

18

10

D5-E

43

60

-21.1

18

11

D5-E

1

18*

-20.4

New, loose profile taped

21

2

A14-O

6

57

-23.1

Replaced fiber

22

3

A14-E

25

76

-20.9

Reseated profile

34

5

B14-E

26

116

-20.1

Replaced fiber

34

7

B14-E

26

116

-21.5

37

10

D6-O

38

120

-24.3

New, reseated profile

37

10

D6-O

39

121

-22.5

New, reseated profile

38

8

D6-E

72

154

-23.6

probably tile, no packs with better light available

38

10

D6-E

38

120

-21.3

New, reseated profile

42

2

A16-E

8

117

-25.6

Reseated profile

44

5

B15-E

27

149

-20.7

New reseated profile

 

QC 6 VERSION 5.8: 10-Sep-01 16:35:58

LIST OF FIBERS/TILES WITH RESPONSE < 75%

FOR SCAN C of ANL037

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

             

3

11

D4-O

15

15

-39.6

Not fiber, not tile, not fixed

6

7

C10-E

1

12

-28.7

Special tiles

6

8

C10-E

1

12

-27.2