QC 6 VERSION 6.0: 7-Nov-01 10:03:59
LIST OF FIBERS/TILES WITH RESPONSE < 80%
FOR ANL039 SCAN A
PMT |
LAYER |
CELL |
TILE |
SLOT |
%DEV |
SOLUTION |
3 |
10 |
D4-O |
1 |
1 |
-20.7 |
Special Tile |
3 |
11 |
D4-O |
1 |
1 |
-82.2 |
Loose special tile - put washer under nut and taped tile to spacer plates |
4 |
10 |
D4-E |
1 |
1 |
-21.7 |
Special tiles |
4 |
11 |
D4-E |
1 |
1 |
-23.5 |
|
5 |
7 |
C10-O |
1 |
12 |
MISSING |
Replaced special tile |
6 |
7 |
C10-E |
1 |
12 |
MISSING |
|
5 |
9 |
C10-O |
1 |
12 |
-21.2 |
Special tiles |
6 |
9 |
C10-E |
1 |
12 |
-25.5 |
|
8 |
2 |
A12-E |
8 |
25 |
-39.3 |
Loose profile reseated |
12 |
1 |
A13-E |
4 |
30 |
-21.4 |
Replaced fiber |
12 |
3 |
A13-E |
4 |
30 |
-24.5 |
|
17 |
8 |
D5-O |
57 |
74 |
-26.9 |
Tile replaced |
18 |
8 |
D5-E |
57 |
74 |
-27.7 |
|
17 |
11 |
D5-O |
65 |
82* |
-45.0 |
Loose profile - shim & reseated |
18 |
8 |
D5-E |
2 |
19 |
MISSING |
Replaced fiber |
18 |
10 |
D5-E |
2 |
19 |
MISSING |
|
18 |
9 |
D5-E |
1 |
18* |
-46.8 |
On side of tile- fixed |
18 |
9 |
D5-E |
50 |
67 |
-31.1 |
Replaced fiber |
18 |
11 |
D5-E |
50 |
67 |
-37.1 |
|
30 |
3 |
A15-E |
13 |
92 |
-22.5 |
Wide slot -Shim tile |
33 |
4 |
B14-O |
17 |
107 |
-33.6 |
Replace fiber |
33 |
6 |
B14-O |
17 |
107 |
-40.3 |
|
34 |
7 |
B14-E |
17 |
107 |
-29.1 |
Remove paint in slot, reseated profile |
34 |
7 |
B14-E |
19 |
109 |
-22.1 |
Reseated profile |
37 |
9 |
D6-O |
47 |
129 |
-24.6 |
Replace fiber |
37 |
11 |
D6-O |
47 |
129 |
-23.6 |
|
41 |
3 |
A16-O |
17 |
126 |
-25.0 |
Wide slot - shim tile |
42 |
2 |
A16-E |
48 |
157 |
-63.1 |
Replaced fiber |
QC 6 VERSION 6.0: 8-Nov-01 16:33:08
LIST OF FIBERS/TILES WITH RESPONSE < 80%
FOR ANL039 SCAN B
PMT |
LAYER |
CELL |
TILE |
SLOT |
%DEV |
SOLUTION |
3 |
10 |
D4-O |
1 |
1 |
-24.1 |
Special Tiles |
4 |
10 |
D4-E |
1 |
1 |
-21.4 |
|
5 |
7 |
C10-O |
1 |
12 |
-32.2 |
|
5 |
9 |
C10-O |
1 |
12 |
-23.4 |
|
6 |
7 |
C10-E |
1 |
12 |
-31.7 |
|
6 |
9 |
C10-E |
1 |
12 |
-24.5 |
|
15 |
7 |
B12-O |
11 |
44 |
-23.4 |
Partly on side of tile - fixed |
17 |
8 |
D5-O |
58 |
75 |
-20.5 |
Reseated profile |
18 |
9 |
D5-E |
50 |
67 |
-39.2 |
Replaced fiber - AGAIN |
18 |
11 |
D5-E |
50 |
67 |
-41.3 |
QC 6 VERSION 6.0: 14-Nov-01 17:16:52
LIST OF FIBERS/TILES WITH RESPONSE < 75%
FOR ANL039 SCAN C
PMT |
LAYER |
CELL |
TILE |
SLOT |
%DEV |
SOLUTION |
5 |
7 |
C10-O |
1 |
12 |
-32.2 |
Special Tile |
6 |
7 |
C10-E |
1 |
12 |
-31.7 |