QC 6 VERSION 6.0: 30-Nov-01 16:51:16

LIST OF FIBERS/TILES WITH RESPONSE < 80%

FOR ANL041- Scan A

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

3

11

D4-O

1

1

-55.6

Special tile - not sitting flat on surface - put washer under rod, taped top and bottom edges

4

11

D4-E

1

1

-51.9

5

7

C10-O

1

12

-27.1

Special tiles

5

9

C10-O

1

12

-22.7

6

9

C10-E

1

12

-32.3

8

2

A12-E

9

26

-78.7

Fiber damaged next to girder - replaced

10

5

B11-E

8

25

-23.1

These 3 fibers all have low (85%) response in other layer - all replaced

11

3

A13-O

1

27

-20.9

11

3

A13-O

2

28

-20.5

15

7

B12-O

18

51

-54.7

Loose profile - shim tile & reseat

17

8

D5-O

1

18*

-28.9

Replaced these 5 tiles and the last one in same layer in C10 cell - new tiles from pack 491: We have no packs that give better light yield

17

8

D5-O

2

19

-28.2

17

8

D5-O

3

20

-32.1

17

8

D5-O

4

21

-24.4

17

8

D5-O

5

22

-25.2

18

8

D5-E

1

18*

-25.1

18

8

D5-E

2

19

-26.4

18

8

D5-E

3

20

-26.6

18

8

D5-E

4

21

-22.9

18

8

D5-E

5

22

-24.6

18

9

D5-E

29

46

-20.3

Fiber replace

18

11

D5-E

29

46

-22.7

18

11

D5-E

42

59

-35.7

Wide slot, fiber not centered - shim tile & reseat

24

4

B13-E

2

62

-36.6

Fiber replaced

24

6

B13-E

2

62

-34.4

24

5

B13-E

27

87

-23.7

Bad paint - clean slot and reseat

29

1

A15-O

25

104

-41.2

Replaced fiber - damaged inside of profile

29

3

A15-O

25

104

-39.0

30

3

A15-E

7

86

-51.6

Fiber not centered on tile - shim tile

33

5

B14-O

14

104

-20.5

Tile not centered - fixed

33

7

B14-O

11

101

-22.1

Shim tile and reseat profile

33

7

B14-O

16

106

-38.9

Profile loose, tile not centered - fixed

34

5

B14-E

1

91

-27.1

Fiber replaced

34

7

B14-E

1

91

-25.4

38

11

D6-E

8

90

-20.1

Paint in slot - cleaned and reseat profile

41

1

A16-O

21

130*

-24.8

Gap shimmed

41

2

A16-O

6

115

-58.5

Fiber on side of tile

41

2

A16-O

11

120

-62.5

Fiber on side of tile

41

3

A16-O

3

112

-20.2

Reseated profile

42

1

A16-E

3

112

-27.6

Fiber replaced

42

3

A16-E

3

112

-25.9

42

1

A16-E

23

132

-51.7

Fiber replaced

42

3

A16-E

23

132

-47.8

42

1

A16-E

48

157

-54.8

Fiber replaced - cladding damage in girder

42

3

A16-E

48

157

-53.6

42

3

A16-E

27

136

-59.0

Fiber not centered on tile - shim tile

43

5

B15-O

15

137

-38.8

Fiber on side of tile, damaged- replaced

44

7

B15-E

15

137

-25.4

Reseated profile

 

QC 6 VERSION 6.0: 6-Dec-01 15:29:11

LIST OF FIBERS/TILES WITH RESPONSE < 80%

FOR ANL041 Scan -B

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

3

11

D4-O

1

1

-44.2

Special Tiles

4

10

D4-E

1

1

-20.6

4

11

D4-E

1

1

-46.0

5

9

C10-O

1

12

-23.5

6

9

C10-E

1

12

-35.1

8

3

A12-E

8

25

-22.5

New-Narrow slot-reseated profile

10

5

B11-E

8

25

-21.1

Fiber replaced - second time

10

7

B11-E

8

25

-18.6

17

8

D5-O

3

20

-25.3

Tiles - no better tiles available

17

8

D5-O

10

27

-20.5

18

11

D5-E

49

66*

-20.1

Gap shimmed

30

1

A15-E

19

98*

-20.2

Gap shimmed

33

5

B14-O

14

104

-20.2

Marginal fiber - not replaced

33

7

B14-O

14

104

-17.3

33

7

B14-O

12

102

-27.5

New - fiber partly on side of tile -fixed

34

5

B14-E

1

91

-37.5

Fiber replaced - second time

34

7

B14-E

1

91

-27.5

38

11

D6-E

10

92

-21.0

Reseated profile

41

1

A16-O

3

112

-20.4

Reseated profile

Also found 2 fibers in A12-Odd side, tiles 7 and 9 in layers 1 and 3 that had been tied too tightly to the rail and were bent - these were replaced.

 

QC 6 VERSION 6.0: 11-Dec-01 16:04:01

LIST OF FIBERS/TILES WITH RESPONSE < 75%

FOR ANL041 Scan C

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

3

11

D4-O

1

1

-26.0

Special Tiles

4

11

D4-E

1

1

-27.4

6

9

C10-E

1

12

-35.1