QC 6 VERSION 6.0: 30-Nov-01 16:51:16
LIST OF FIBERS/TILES WITH RESPONSE < 80%
FOR ANL041- Scan A
PMT |
LAYER |
CELL |
TILE |
SLOT |
%DEV |
SOLUTION |
3 |
11 |
D4-O |
1 |
1 |
-55.6 |
Special tile - not sitting flat on surface - put washer under rod, taped top and bottom edges |
4 |
11 |
D4-E |
1 |
1 |
-51.9 |
|
5 |
7 |
C10-O |
1 |
12 |
-27.1 |
Special tiles |
5 |
9 |
C10-O |
1 |
12 |
-22.7 |
|
6 |
9 |
C10-E |
1 |
12 |
-32.3 |
|
8 |
2 |
A12-E |
9 |
26 |
-78.7 |
Fiber damaged next to girder - replaced |
10 |
5 |
B11-E |
8 |
25 |
-23.1 |
These 3 fibers all have low (85%) response in other layer - all replaced |
11 |
3 |
A13-O |
1 |
27 |
-20.9 |
|
11 |
3 |
A13-O |
2 |
28 |
-20.5 |
|
15 |
7 |
B12-O |
18 |
51 |
-54.7 |
Loose profile - shim tile & reseat |
17 |
8 |
D5-O |
1 |
18* |
-28.9 |
Replaced these 5 tiles and the last one in same layer in C10 cell - new tiles from pack 491: We have no packs that give better light yield |
17 |
8 |
D5-O |
2 |
19 |
-28.2 |
|
17 |
8 |
D5-O |
3 |
20 |
-32.1 |
|
17 |
8 |
D5-O |
4 |
21 |
-24.4 |
|
17 |
8 |
D5-O |
5 |
22 |
-25.2 |
|
18 |
8 |
D5-E |
1 |
18* |
-25.1 |
|
18 |
8 |
D5-E |
2 |
19 |
-26.4 |
|
18 |
8 |
D5-E |
3 |
20 |
-26.6 |
|
18 |
8 |
D5-E |
4 |
21 |
-22.9 |
|
18 |
8 |
D5-E |
5 |
22 |
-24.6 |
|
18 |
9 |
D5-E |
29 |
46 |
-20.3 |
Fiber replace |
18 |
11 |
D5-E |
29 |
46 |
-22.7 |
|
18 |
11 |
D5-E |
42 |
59 |
-35.7 |
Wide slot, fiber not centered - shim tile & reseat |
24 |
4 |
B13-E |
2 |
62 |
-36.6 |
Fiber replaced |
24 |
6 |
B13-E |
2 |
62 |
-34.4 |
|
24 |
5 |
B13-E |
27 |
87 |
-23.7 |
Bad paint - clean slot and reseat |
29 |
1 |
A15-O |
25 |
104 |
-41.2 |
Replaced fiber - damaged inside of profile |
29 |
3 |
A15-O |
25 |
104 |
-39.0 |
|
30 |
3 |
A15-E |
7 |
86 |
-51.6 |
Fiber not centered on tile - shim tile |
33 |
5 |
B14-O |
14 |
104 |
-20.5 |
Tile not centered - fixed |
33 |
7 |
B14-O |
11 |
101 |
-22.1 |
Shim tile and reseat profile |
33 |
7 |
B14-O |
16 |
106 |
-38.9 |
Profile loose, tile not centered - fixed |
34 |
5 |
B14-E |
1 |
91 |
-27.1 |
Fiber replaced |
34 |
7 |
B14-E |
1 |
91 |
-25.4 |
|
38 |
11 |
D6-E |
8 |
90 |
-20.1 |
Paint in slot - cleaned and reseat profile |
41 |
1 |
A16-O |
21 |
130* |
-24.8 |
Gap shimmed |
41 |
2 |
A16-O |
6 |
115 |
-58.5 |
Fiber on side of tile |
41 |
2 |
A16-O |
11 |
120 |
-62.5 |
Fiber on side of tile |
41 |
3 |
A16-O |
3 |
112 |
-20.2 |
Reseated profile |
42 |
1 |
A16-E |
3 |
112 |
-27.6 |
Fiber replaced |
42 |
3 |
A16-E |
3 |
112 |
-25.9 |
|
42 |
1 |
A16-E |
23 |
132 |
-51.7 |
Fiber replaced |
42 |
3 |
A16-E |
23 |
132 |
-47.8 |
|
42 |
1 |
A16-E |
48 |
157 |
-54.8 |
Fiber replaced - cladding damage in girder |
42 |
3 |
A16-E |
48 |
157 |
-53.6 |
|
42 |
3 |
A16-E |
27 |
136 |
-59.0 |
Fiber not centered on tile - shim tile |
43 |
5 |
B15-O |
15 |
137 |
-38.8 |
Fiber on side of tile, damaged- replaced |
44 |
7 |
B15-E |
15 |
137 |
-25.4 |
Reseated profile |
QC 6 VERSION 6.0: 6-Dec-01 15:29:11
LIST OF FIBERS/TILES WITH RESPONSE < 80%
FOR ANL041 Scan -B
PMT |
LAYER |
CELL |
TILE |
SLOT |
%DEV |
SOLUTION |
3 |
11 |
D4-O |
1 |
1 |
-44.2 |
Special Tiles |
4 |
10 |
D4-E |
1 |
1 |
-20.6 |
|
4 |
11 |
D4-E |
1 |
1 |
-46.0 |
|
5 |
9 |
C10-O |
1 |
12 |
-23.5 |
|
6 |
9 |
C10-E |
1 |
12 |
-35.1 |
|
8 |
3 |
A12-E |
8 |
25 |
-22.5 |
New-Narrow slot-reseated profile |
10 |
5 |
B11-E |
8 |
25 |
-21.1 |
Fiber replaced - second time |
10 |
7 |
B11-E |
8 |
25 |
-18.6 |
|
17 |
8 |
D5-O |
3 |
20 |
-25.3 |
Tiles - no better tiles available |
17 |
8 |
D5-O |
10 |
27 |
-20.5 |
|
18 |
11 |
D5-E |
49 |
66* |
-20.1 |
Gap shimmed |
30 |
1 |
A15-E |
19 |
98* |
-20.2 |
Gap shimmed |
33 |
5 |
B14-O |
14 |
104 |
-20.2 |
Marginal fiber - not replaced |
33 |
7 |
B14-O |
14 |
104 |
-17.3 |
|
33 |
7 |
B14-O |
12 |
102 |
-27.5 |
New - fiber partly on side of tile -fixed |
34 |
5 |
B14-E |
1 |
91 |
-37.5 |
Fiber replaced - second time |
34 |
7 |
B14-E |
1 |
91 |
-27.5 |
|
38 |
11 |
D6-E |
10 |
92 |
-21.0 |
Reseated profile |
41 |
1 |
A16-O |
3 |
112 |
-20.4 |
Reseated profile |
Also found 2 fibers in A12-Odd side, tiles 7 and 9 in layers 1 and 3 that had been tied too tightly to the rail and were bent - these were replaced.
QC 6 VERSION 6.0: 11-Dec-01 16:04:01
LIST OF FIBERS/TILES WITH RESPONSE < 75%
FOR ANL041 Scan C
PMT |
LAYER |
CELL |
TILE |
SLOT |
%DEV |
SOLUTION |
3 |
11 |
D4-O |
1 |
1 |
-26.0 |
Special Tiles |
4 |
11 |
D4-E |
1 |
1 |
-27.4 |
|
6 |
9 |
C10-E |
1 |
12 |
-35.1 |