QC 6 VERSION 6.0: 27-Mar-02 12:16:37

LIST OF FIBERS/TILES WITH RESPONSE < 80%

FOR ANL047 SCAN A

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

3

10

D4-O

2

2

-32.8

Fiber replaced

4

11

D4-E

4

4

-22.9

Reseated profile

4

11

D4-E

7

7

-24.1

Reseated profile

4

11

D4-E

12

12

-40.8

Fiber replaced

5

7

C10-O

1

12

-34.2

Special tiles

5

8

C10-O

1

12

-27.6

6

7

C10-E

1

12

-29.9

6

8

C10-E

1

12

-38.7

6

9

C10-E

1

12

-68.1

Loose fiber on special tile

8

2

A12-E

3

20

-22.7

Fiber replaced

9

5

B11-O

14

31

-20.5

Shim tile - reseat profile

9

7

B11-O

7

24

-39.5

On side of tile - fixed

12

1

A13-E

6

32

-23.1

Fiber replaced

12

3

A13-E

6

32

-26.0

15

4

B12-O

11

44

-28.3

Fiber replaced

15

6

B12-O

11

44

-21.7

16

7

B12-E

1

34*

-26.4

Shim tile - reseat profile

16

7

B12-E

3

36

-40.8

Profile loose - reseated

18

9

D5-E

46

63

-37.4

Fiber replaced

18

11

D5-E

46

63

-28.2

18

9

D5-E

50

67

-25.3

Reseat Profile

18

11

D5-E

50

67

-16.4

18

11

D5-E

4

21

-24.9

Reseat Profile

18

11

D5-E

17

34*

-31.3

Profile loose - reseated

18

11

D5-E

18

35

-88.2

On side of tile - fixed

18

11

D5-E

24

41

-25.2

Reseat Profile

18

11

D5-E

33

50*

-28.1

Loose profile -Reseat

18

11

D5-E

34

51

-65.4

On side of tile - fixed

21

1

A14-O

15

66*

-32.2

Fiber replaced

21

3

A14-O

15

66*

-33.2

21

2

A14-O

5

56

-46.3

Fiber replaced

21

3

A14-O

25

76

-27.2

Fiber not centered on tile

22

1

A14-E

4

55

-24.7

Reseated profile

22

3

A14-E

3

54

-41.5

Loose fiber - fixed

33

7

B14-O

29

119

-21.6

Reseated profile

34

4

B14-E

17

107

-47.2

Fiber replaced

34

6

B14-E

17

107

-47.2

34

5

B14-E

27

117

-51.6

Fiber replaced

34

7

B14-E

27

117

-52.5

37

9

D6-O

51

133

-20.8

Reseated profile

37

10

D6-O

62

144

-28.9

Reseated profile

37

11

D6-O

48

130*

-54.5

On side of tile - fixed

38

11

D6-E

6

88

MISSING

No problem found - reseated profile

38

11

D6-E

48

130*

-88.8

On side of tile - fixed

38

11

D6-E

50

132

-79.2

On side of tile - fixed

41

1

A16-O

5

114*

-32.1

Replaced tile with one from pack 867

42

1

A16-E

5

114*

-30.6

42

1

A16-E

41

150

-30.8

Fiber replaced

42

3

A16-E

41

150

-34.5

42

3

A16-E

25

134

-24.4

Shim tile & reseat profile

43

4

B15-O

32

154

-27.5

Fiber replaced

43

6

B15-O

32

154

-28.3

43

5

B15-O

13

135

-34.1

Loose profile, reseated

43

5

B15-O

32

154

MISSING

Fiber glued in wrong location - replace it

43

7

B15-O

10

132

-29.6

Reseated profile

43

7

B15-O

12

134

-20.9

Reseated profile

QC 6 VERSION 6.0: 1-Apr-02 13:27:06

LIST OF FIBERS/TILES WITH RESPONSE < 80%

FOR ANL047 SCAN B

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

5

7

C10-O

1

12

-32.1

Special tiles

5

8

C10-O

1

12

-25.5

6

7

C10-E

1

12

-30.0

6

8

C10-E

1

12

-33.3

6

9

C10-E

1

12

-77.1

Fiber loose on special tile - fixed again

8

2

A12-E

7

24

-20.9

Reseated profile

9

7

B11-O

15

32

-20.5

Reseated profile

18

8

D5-E

44

61

-21.7

Reseated profile

18

10

D5-E

44

61

-16.5

Reseated profile

18

9

D5-E

50

67

-20.2

Reseated profile

18

11

D5-E

50

67

-21.8

21

1

A14-O

25

76

-21.3

Fiber replaced

21

3

A14-O

25

76

-29.2

34

5

B14-E

27

117

-21.4

Reseated profile

34

7

B14-E

27

117

-20.1

37

10

D6-O

13

95

-25.6

Cleaned tile, Reseated profile

42

1

A16-E

25

134

-22.2

Reseated profile

42

3

A16-E

25

134

-22.9

QC 6 VERSION 6.0: 5-Apr-02 14:09:56

LIST OF FIBERS/TILES WITH RESPONSE < 75%

FOR ANL047 SCAN C

PMT

LAYER

CELL

TILE

SLOT

%DEV

SOLUTION

5

7

C10-O

1

12

-33.6

Special Tiles

6

7

C10-E

1

12

-25.5

6

8

C10-E

1

12

-32.2